- Response Time: <3 sec
- Lowest Detection Limit: 0.5 nM
- Sensitivity: 1 pA/nM
- Application: Microvessels
- Requires cable #91580
7 micron sensor with exceptional performance
The ISO-NOP007 has a tip diameter of just 7 microns and a length of 2mm. The response is linear over a wide dynamic concentration range of NO. The design is based on a single carbon fiber coated with WPI’s NO-selective membrane. A detection limit of approximately 1 nM NO makes this electrode ideal for use in tissues and microvessels.
The response of a 7μm NO sensor (ISO-NOP007) to successive additions of NO (100nM). Inset shows the linearity of the resulting calibration plot.
Simultaneous measurement of force (top trace) and changes of NO concentration (lower trace) in (A) the rat superior mesenteric artery relaxed with ACh and (B) a small human artery relaxed with ACh and SNAP. In this artery oxyhaemoglobin (oxyHb) partly reversed the increase in NO concentration with only a small change in force. [U. Simonsen, et al., J. Physiol., 1999, 516: 271-282.]
| Outside Diameter | 7μm |
| Available Length | 2mm |
| Response Time | < 3 seconds |
| Lowest Detection Limit/Range | 0.5nM |
| Nominal Sensitivity-New sensor | ≥1pA/nM |
| Baseline Drift | none |
| Poise Voltage | 865mV |
| Typical Quiescent Baseline Current, 25°C | 300pA |
| Acceptable Baseline Range | 200-1500pA |
| Polarization Time | 1+ hours |
| 7363-4 | Enzol Enzymatic Detergent, 1 gallon |
| 7364 | Cidex Disinfecting Solution |
| 91580 | Microsensor Cable |
| 47510 | ProGuide Position/Holder with Base |
| 47520 | ProGuide Position/Holder |
| 47530 | ProGuide Pluse Position/Holder with fina adjustment |
| 47540 | ProGuide Plus with Base |
| SNAP25 | SNAP, 25mg vial |
| SNAP50 | SNAP, 50mg vial |
| SNAP100 | SNAP, 100mg vial |